For laboratory measurements, with few exceptions all probe types can be used for all sample materials in the measuring range of 0.1 to 10 W/mK. In most cases you can choose a probe by shape and size of the available sample pieces or by sample preparation possibilities. For in-situ tests always the Field VLQ should be used, the smaller laboratory probes are not adequate for this purpose.

We offer complete TK04 measuring kits (VLQ kit, HLQ kit or Complete kit with HLQ and VLQ probe), including all necessary hardware and software components. But all probes are available separately for replacing defect probes or for upgrading an existing TK04 system.

Standard VLQ

Standard VLQ:  needle probe for TK04 thermal conductivity meter

Standard VLQ 70 / 60 for laboratory use
(choose length according to sample size)
Probe type needle probe
Measuring range 0.1 - 10 W/mK
Accuracy ± 2 %
Duration of 1 measurement 80 s
Diameter (needle) 2 mm
Length (needle) 70 mm (VLQ 70) / 60 mm (VLQ 60)
Dimensions (overall) length:160 mm (VLQ 70) / 150 mm (VLQ 60), Ø: 16 mm (max)
Minimum sample size Ø: 40 mm *, length: 85mm (VLQ 70) / 75 mm (VLQ 60) *
Sample shape any
Maximum sample size no limit
Sample temperature -25 to 125°C
Supplied parts probe, INI file, carrying case, protective cap, drill bits, data sheet

* Absolute minimum size, valid for low thermal conductivity materials. For higher thermal conductivities, larger samples are required.

Technical specifications are subject to change without prior notice.

Mini VLQ

Mini VLQ:  small needle probe for TK04 thermal conductivity meter

Mini VLQ for laboratory use
and small samples
Probe type small needle probe
Measuring range 0.3 - 6 W/mK
Accuracy ± 2 %
Duration of 1 measurement 80 s
Diameter (needle) 2 mm
Length (needle) 50 mm
Dimensions (overall) length:140 mm, Ø: 16 mm (max)
Minimum sample size Ø: 40 mm *, length: 60mm *
Sample shape any
Maximum sample size no limit
Sample temperature -25 to 125°C
Supplied parts probe, INI file, carrying case, protective cap, drill bits, data sheet

* Absolute minimum size, valid for low thermal conductivity materials. For higher thermal conductivities, larger samples are required.

Technical specifications are subject to change without prior notice.

Field VLQ

Field VLQ: large needle probe for TK04 thermal conductivity meter

Field VLQ for field use
Probe type large needle probe
Measuring range 0.6 - 4 W/mK
Accuracy ± 5 %
Duration of 1 measurement 240 s
Diameter (needle) 6 mm
Length (needle) 300 mm
Dimensions (overall) length: 420 mm, Ø: 42 mm (max, w/o connector)
Minimum sample size  
Sample shape any
Maximum sample size no limit
Sample temperature -25 to 50°C
Supplied parts probe, INI file, carrying case, protective cap, 5 guiding tubes, insertion tool, data sheet

Technical specifications are subject to change without prior notice.

Standard HLQ

Standard HLQ: probe for plane surfaces for TK04 thermal conductivity meter

Standard HLQ for laboratory use
Probe type probe for plane surfaces
Measuring range 0.3 - 10 W/mK
Accuracy ± 2 %
Duration of 1 measurement 80 s
Diameter (needle) 2 mm
Length (needle) 70 mm
Dimensions (overall) Ø: 88 mm, height: 30 mm (w/o connector)
Minimum sample size Ø: 90 mm *, height: 20 mm *
Sample shape any
Maximum sample size no limit **
Sample temperature -25 to 70°C
Supplied parts probe, INI file, carrying case, data sheet

* Absolute minimum size, valid for low thermal conductivity materials. For higher thermal conductivities, larger samples are required.

** With pressure device: max. height 130 mm.

Technical specifications are subject to change without prior notice.

Mini HLQ

Mini HLQ: small probe for plane surfaces for TK04 thermal conductivity meter

Mini HLQ for laboratory use
and small samples
Probe type small probe for plane surfaces
Measuring range 0.3 - 3 W/mK
Accuracy ± 5 %
Duration of 1 measurement 60 s
Diameter (needle) 2.0 mm
Length (needle) 45 mm
Dimensions (overall) Ø: 50 mm, height: 30 mm (w/o connector)
Minimum sample size Ø: 50 mm *, height: 20 mm *
Sample shape any
Maximum sample size no limit **
Sample temperature -25 to 70°C
Supplied parts probe, INI file, carrying case, data sheet

* Absolute minimum size, valid for low thermal conductivity materials. For higher thermal conductivities, larger samples are required.

Technical specifications are subject to change without prior notice.

 

Options

 

General

Optional accessories are available to complete the TK04 measuring setup for particular sample types and measuring tasks.

When using a Standard or Mini HLQ the use of a lever press is recommended to ensure good contact between probe and sample by applying controlled contact pressure.

A sample container matching the diameter of the Standard HLQ probe is required for testing liquids, pastes, powder or fragments.

Guiding tubes provide fixed measuring positions for safely insert needle probes into hard samples or in the field.

Additional reference materials in different sizes serve to check the operating condition of instrument and probes.

Lever press

Lever press with pressure limiter

Lever press with pressure limiter
Application ensures good contact between probes for plane surfaces (Standard HLQ, Mini HLQ) and samples by applying controlled pressure
Contact pressure 0 to 10 bar/1.7 kN, stepless
Dimensions approx. 290 x 150 mm, height 420/575 mm (lever down/up)
Weight 12.9 kg
Sample size height max. 13 cm, other dimensions practically unlimited
Supplied parts lever press, screws, data sheet (delivered w/o sample and probe)

Technical specifications are subject to change without prior notice.

Sample container

Sample container for liquids, pastes, powder and fragments

Sample container   
Application testing liquids, pastes, powder or fragments with the Standard HLQ
Dimensions Ø 108 mm, height 81 mm (outer dimensions), inner diamater matching the Standard HLQ
Weight 1800 g
Supplied parts sample container, 2 handles for removing the probee

Technical specifications are subject to change without prior notice.

Guiding tubes

 

Guiding tubes for Standard VLQ

Guiding tubes for Standard VLQ

Guiding tubes for Field VLQ

Guiding tubes for Field VLQ

Application fixed measuring position for safe insertion of needle probes (Standard VLQ) into hardening samples (re-usable) fixed measuring position for safe insertion of large needle probes (Field VLQ) in the field (re-usable)
Material stainless steel V2A stainless steel V2A
Outer diameter 2.5 mm 7 mm
Length ca. 75 mm ca. 320 mm
Supplied parts 10 guiding tubes 10 guiding tubes, insertion tool, case

Technical specifications are subject to change without prior notice.

Reference materials

 

Reference material ST 1.6 for Standard VLQ

ST 1.6 for Standard VLQ

Reference material ST 1.6 for Standard HLQ

ST 1.6 for Standard HLQ

Thermal conductivity * 1.6 W/mK 1.6 W/mK
Dimensions ** 55 x 55 x 90 mm 110 x 90 x 30 mm
Drill hole Ø 2 mm none
Physical properties download datasheet download datasheet
Supplied parts Reference material, TC certificate Reference material, TC certificate
Application Checking the operating condition of instrument and probes. The user is not required to perform any calibration or reference measurements! 
 

 

Combined Reference material ST 1.6 for Standard VLQ and HLQ

ST 1.6 for Standard Probes

 

Reference material ST 0.2 for Standard VLQ

ST 0.2 for Standard VLQ

Thermal conductivity * 1.6 W/mK 0.2 W/mK
Dimensions ** 110 x 90 x 55 mm Ø 60 mm, height 90 mm
Drill hole Ø 2 mm Ø 2 mm
Physical properties download datasheet download datasheet
Supplied parts Reference material, TC certificate Reference material, TC certificate
Application Checking the operating condition of instrument and probes. The user is not required to perform any calibration or reference measurements! 

* The specified thermal conductivities are approximate values. Each reference block is calibrated individually before delivery and is labeled with its exact thermal conductivity value.

** Dimensions may vary slightly.