Technical specification for the TK04 thermal conductivity meter

Reference standards The specified thermal conductivities are approximate values. Each standard is calibrated individually before delivery and is labeled with its exact thermal conductivity value. Dimensions may vary slightly. A datasheet summarizing physical properties and sizes of the reference standards is available for download.

Reference standard ST 1.6 for full-space probes
Name Standard ST 1.6 (thermal conductivity approx. 1.6 W/mK)
for half-space probe 110 x 90 x 30 mm
for full-space probes 55 x 55 x 90 mm, with drill hole (Ø 2 mm)
for both probe types 110 x 90 x 55 mm, with drill hole (Ø 2 mm)
Reference standard ST 0.2 for full-space probes
Name Standard ST 0.2 (thermal conductivity approx. 0.2 W/mK)
for full-space probes Ø 60 mm, height 90 mm, with drill hole (Ø 2 mm)

The users are not required to perform calibration or reference tests themselves. The standards are intended for occasionally checking the function of instrument and probes.

Subject to change without prior notice. Back to top