Technical specification for the TK04 thermal conductivity meter
- instrument
- probes
- software
- reference standards
- options
- show all
Instrument
![]() |
|
| Name | TK04 measuring unit |
| Measuring principle | transient heat flow (needle probe method) |
| Measuring range | 0.1 to 10 W/mK (probe dependent, see below) |
| Accuracy | ± 2 % (probe dependent, see below) |
| Reproducibility | ± 1.5 % |
| Heater current precision | ± 0.01 % |
| Duration of 1 measurement | typically 80 sec (probe dependent, see below) |
| Automatic repetitions | up to 99 measurements per series (unattended) |
| Sample sizes | no upper size limit, arbitrary shape, minimum size probe dependent (see below) |
| Operating temperature | 0 to 45°C |
| Sample temperature | 0 to 70°C (-25 to 125°C on request) |
| Power supply | 220/240 V ~ (50 Hz) ; 100/120 V ~ (60 Hz) |
| Power consumption | ~ 40W |
| Dimensions (W x H x D) | 471 x 147 x 391 mm |
| Weight | 13 kg |
| Interface | RS232 serial interface |
Subject to change without prior notice. Back to top
Probes The minimum sample size indicates the minimal dimensions required for the particular probe type. For testing of materials with high thermal conductivity values larger samples my be necessary.
![]() |
|
| Name | Standard VLQ |
| Probe type | full-space for laboratory use |
| Dimensions | length: 70 mm, diameter: 2 mm (without handle) |
| Measuring range | 0.1 to 10 W/mK |
| Accuracy | ± 2 % |
| Duration of 1 measurement | 80 sec |
| Minimum sample size | diameter: approx. 30 mm, length: 75 mm |
![]() |
|
| Name | Standard HLQ |
| Probe type | half-space for laboratory use |
| Dimensions | diameter: 88 mm, height: 30 mm |
| Measuring range | 0.3 to 10 W/mK |
| Accuracy | ± 2 % |
| Duration of 1 measurement | 80 sec |
| Minimum sample size | diameter: approx. 80 mm, height: approx. 15 mm |
![]() |
|
| Name | Mini HLQ |
| Probe type | half-space for laboratory use with small samples |
| Dimensions | diameter: 50 mm, height: 30 mm |
| Measuring range | 0.3 to 3 W/mK |
| Accuracy | ± 5 % |
| Duration of 1 measurement | 60 sec |
| Minimum sample size | diameter: approx. 50 mm, height: approx. 15 mm |
![]() |
|
| Name | Field VLQ |
| Probe type | full-space for in-situ use |
| Dimensions | length: 300 mm, diameter: 6 mm |
| Measuring range (tested) | 0.6 to 4 W/mK |
| Accuracy | ± 5 % |
| Duration of 1 measurement | 240 sec |
Subject to change without prior notice. Back to top
Software With the TK04 multi-user licence you may install the software on any number of computers required. The modules for measuring/evaluation and for graphics/quality analysis can be installed and used independently.
| Operating system | Windows 98, Me, 2000, XP, 7 |
| CPU | 1 GHz or higher |
| RAM | 512 MB or higher |
| Interface | 1 free serial interface (COM-Port RS232) or USB port with USB-to-serial converter |
| Screen resolution | 1280 x 1024 |
Subject to change without prior notice. Back to top
Reference standards The specified thermal conductivities are approximate values. Each standard is calibrated individually before delivery and is labeled with its exact thermal conductivity value. Dimensions may vary slightly. A datasheet summarizing physical properties and sizes of the reference standards is available for download.
![]() |
|
| Name | Standard ST 1.6 (thermal conductivity approx. 1.6 W/mK) |
| for half-space probe | 110 x 90 x 30 mm |
| for full-space probes | 55 x 55 x 90 mm, with drill hole (Ø 2 mm) |
| for both probe types | 110 x 90 x 55 mm, with drill hole (Ø 2 mm) |
![]() |
|
| Name | Standard ST 0.2 (thermal conductivity approx. 0.2 W/mK) |
| for full-space probes | Ø 60 mm, height 90 mm, with drill hole (Ø 2 mm) |
The users are not required to perform calibration or reference tests themselves. The standards are intended for occasionally checking the function of instrument and probes.
Subject to change without prior notice. Back to top
Options
![]() |
|
| Name | Pressure device with limiter |
| Usage | pressure device with limiter (max. 10 bar / 1.7 kN) for applying controlled pressure to a half-space probe placed on top of a sample |
| Contact pressure | 0 to 10 bar/1.7 kN |
| Dimensions | 290 x 150 mm, height: 420 mm (lever down) / 575 mm (lever up) |
| Weight | 12.9 kg |
| Sample size | height max. 13 cm, other dimensions practically unlimited |
![]() |
|
| Name | Sample container |
| Usage | sample container for powder, fragments and liquids |
| Dimensions | diameter: 100 mm, height: 90 mm (external dimensions, inner diameter matching the Standard HLQ half-space probe) |
| Weight | approx. 1800 g |
Subject to change without prior notice. Back to top